Skip to content

Parker Quality & Compliance

PQC Reg Watch

Sponsor this week — one slot, $75, newsletter line and this bar

Device recall

Medline Convenience Kits: 1) C-SECTION PREP, Model Number: DYNJ32702B

FDA recall source

Recall number
Z-3060-2026
Event id
99334
Status
Ongoing
FDA classification
Class II
Recalling firm
Medline Industries, LP
Report date
Sep 9, 2026
Posted
Aug 31, 2026
Initiated
Jun 12, 2026
Center classification date
Aug 31, 2026
Product code
OHM
Device type
Cesarean Section Tray
Reason
The kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach of sterility in the packaging, due to wrinkles in the paper lidding which may extend to the seal area. If the sterile barrier is compromised, patients and users may be exposed to microbial contamination. Potential health consequences may include no injury, localized infection, or, less commonly, more serious infection requiring medical treatment, antimicrobial therapy, procedural intervention, or hospitalization. In rare cases, particularly in critically ill or immunocompromised patients, severe infection could become life-threatening.
Root cause
Nonconforming Material/Component
Quantity
352 kits
Code info
1) DYNJ32702B, UDI-DI: 10193489431070(each), 40193489431071(case), Lot Number: 25BBG443; 2) DYNJ32702B, UDI-DI: 10193489431070(each), 40193489431071(case), Lot Number: 25EBL829; 3) DYNJ32702B, UDI-DI: 10193489431070(each), 40193489431071(case), Lot Number: 24EBR761; 4) DYNJ32702B, UDI-DI: 10193489431070(each), 40193489431071(case), Lot Number: 24GBC622; 5) DYNJ32702B, UDI-DI: 10193489431070(each), 40193489431071(case), Lot Number: 24GBX458; 6) DYNJ32702B, UDI-DI: 10193489431070(each), 40193489431071(case), Lot Number: 23LBT973
Distribution
US Nationwide distribution.
Voluntary or mandated
Voluntary: Firm initiated
Location
Northfield, IL, United States