Device recall
LUX-Dx, LUX-Dx II, LUX-Dx II+ Insertable Cardiac Monitor System, SERVER SW LAT CLARITY DS ANZ, Model Number 7248, All Versions (Current 7.7)
- Recall number
- Z-3083-2026
- Event id
- 99524
- Status
- Ongoing
- FDA classification
- Class II
- Recalling firm
- Boston Scientific Corporation
- Report date
- Sep 9, 2026
- Posted
- Sep 2, 2026
- Initiated
- Aug 6, 2026
- Center classification date
- Sep 2, 2026
- Product code
- MXD
- Device type
- Recorder, Event, Implantable Cardiac, (With Arrhythmia Detection)
- Reason
- There is potential for a delay in AF Burden Alert notification affecting the server software which supports the Insertable Cardiac Monitors (ICMs). Delayed notification of an AF Burden Alert may delay clinical evaluation or therapeutic intervention of an atrial arrhythmia.
- Root cause
- Software design
- Quantity
- N/A
- Code info
- UDI-DI: 00802526613852
- Distribution
- Worldwide - US Nationwide distribution.
- Voluntary or mandated
- Voluntary: Firm initiated
- Related 510(k)s
- K231328
- Location
- Saint Paul, MN, United States