Skip to content

Parker Quality & Compliance

PQC Reg Watch

Sponsor this week — one slot, $75, newsletter line and this bar

Device recall

5008X HD/HDF with CLiC Blood Tubing Set Part number 03-5300-3C

FDA recall source

Recall number
Z-3086-2026
Event id
99556
Status
Ongoing
FDA classification
Class II
Recalling firm
Fresenius Medical Care Renal Therapies Group, LLC
Report date
Sep 9, 2026
Posted
Sep 2, 2026
Initiated
Aug 10, 2026
Center classification date
Sep 2, 2026
Product code
KOC
Device type
Accessories, Blood Circuit, Hemodialysis
Reason
Heightened probability of blood leak from the connection between the Arterial Alpha Clip and the main line or pump segment. Leak during treatment may lead to blood loss and hypovolemia.
Root cause
Component change control
Quantity
18,600
Code info
Lots 25JR01801 25KR01225 25KR01251 25NR01016 25NR01265 UDI-DI 00840861102532
Distribution
Worldwide - US Nationwide distribution in the states of AL, CA, CO, FL, GA, IL, KY, LA, MA, MD, MN, MO, NC, NH, NH, NY, OH, PA, TN, TX, VA and the country of Lithuania.
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K243505
Location
Waltham, MA, United States