Device recall
5008X HD/HDF with CLiC Blood Tubing Set Part number 03-5300-3C
- Recall number
- Z-3086-2026
- Event id
- 99556
- Status
- Ongoing
- FDA classification
- Class II
- Recalling firm
- Fresenius Medical Care Renal Therapies Group, LLC
- Report date
- Sep 9, 2026
- Posted
- Sep 2, 2026
- Initiated
- Aug 10, 2026
- Center classification date
- Sep 2, 2026
- Product code
- KOC
- Device type
- Accessories, Blood Circuit, Hemodialysis
- Reason
- Heightened probability of blood leak from the connection between the Arterial Alpha Clip and the main line or pump segment. Leak during treatment may lead to blood loss and hypovolemia.
- Root cause
- Component change control
- Quantity
- 18,600
- Code info
- Lots 25JR01801 25KR01225 25KR01251 25NR01016 25NR01265 UDI-DI 00840861102532
- Distribution
- Worldwide - US Nationwide distribution in the states of AL, CA, CO, FL, GA, IL, KY, LA, MA, MD, MN, MO, NC, NH, NH, NY, OH, PA, TN, TX, VA and the country of Lithuania.
- Voluntary or mandated
- Voluntary: Firm initiated
- Related 510(k)s
- K243505
- Location
- Waltham, MA, United States