Device recall

Bravo CF capsule delivery device, 5-pack, Product Number FGS-0635, Bravo CF capsule delivery device, 1-pack, Product Number FGS-0636. Used to attach a pH monitoring capsule to the esophagus.

FDA recall source

Recall number
Z-3090-2026
Event id
99626
Status
Ongoing
FDA classification
Class I
Recalling firm
Given Imaging Ltd.
Report date
Sep 16, 2026
Posted
Sep 11, 2026
Initiated
Aug 6, 2026
Center classification date
Sep 10, 2026
Product code
FFT
Device type
Electrode, Ph, Stomach
Reason
Capsule may fail to attach to the esophagus or detach from the delivery device. Risks include patient aspiration/inhalation, perforation of esophagus, obstruction of the airway, hemorrhage/blood loss/bleeding, laceration of the esophagus, a delay in diagnosis, and foreign bodies remaining in the patient.
Root cause
Nonconforming Material/Component
Quantity
127,416
Code info
Product Number FGS-0635 UDI-DI (GTIN) 07290101369707 Product Number FGS-0636 UDI-DI (GTIN) 07290101369714 Expiration dates prior to and including October 20, 2027 (lot numbers from 64553F up to and including 68912F)
Distribution
Worldwide distribution. US nationwide including Puerto Rico; Argentina, Australia, Austria, Bahrain, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Czech Republic, Dominican Republic, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Kazakhstan, Kenya, Luxembourg, Netherlands, New Zealand, Norway, Panama, Poland, Portugal, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad And Tobago, United Arab Emirates, and United Kingdom.
Voluntary or mandated
Voluntary: Firm initiated
Location
Yokne'Am Ilit, Israel