Device recall
Presource Packs Containing CliniClean Pericare Wipes within BD Urinary Catheter Trays: 1) KIT,ROBOTICS DH , Catalog Number: PGOCRDNHD; 2) KIT, EAS ROBOTIC , Catalog Number: PG36RPEAB; 3) KIT, VIDEO THORACOSCOPY CMC , Catalog Number: PVCGVTACF
- Recall number
- Z-3097-2026
- Event id
- 99553
- Status
- Ongoing
- FDA classification
- Class II
- Recalling firm
- Cardinal Health 200, LLC
- Report date
- Sep 9, 2026
- Posted
- Sep 3, 2026
- Initiated
- Jul 29, 2026
- Center classification date
- Sep 3, 2026
- Product code
- NWU
- Device type
- Endoscope Introducer Kit
- Reason
- The Urinary Catheter Trays contain CliniClean Pericare Wipes which may potentially be contaminated with a lubricant used during the manufacturing process. If a contaminated wipe is used on a patient, localized skin effects may occur. These may include irritation, erythema, or mild inflammation.
- Root cause
- Process change control
- Quantity
- 115 units
- Code info
- 1) PGOCRDNHD, UDI-DI: 10198956057012(each), 50198956057010(case), Lot Number: NGKV3390 ; 2) PG36RPEAB, UDI-DI: 10198956225091(each), 50198956225099(case), Lot Number: NGKU2052 ; 3) PVCGVTACF, UDI-DI: 10195594816568(each), 50195594816566(case), Lot Number: NGKV3487 ; 4) PVCGVTACF, UDI-DI: 10195594816568(each), 50195594816566(case), Lot Number: NGKV3487
- Distribution
- US Nationwide distribution.
- Voluntary or mandated
- Voluntary: Firm initiated
- Location
- Waukegan, IL, United States