Device recall
Presource Packs Containing CliniClean Pericare Wipes within BD Urinary Catheter Trays: 1) PBCGLVBLI, Catalog Number: KIT, BLO LAVH ; 2) PB15HYFCY, Catalog Number: KIT, UMMC AB HYST II 170032 ; 3) SMA4FTLSCA, Catalog Number: TLH PACK
- Recall number
- Z-3103-2026
- Event id
- 99553
- Status
- Ongoing
- FDA classification
- Class II
- Recalling firm
- Cardinal Health 200, LLC
- Report date
- Sep 9, 2026
- Posted
- Sep 3, 2026
- Initiated
- Jul 29, 2026
- Center classification date
- Sep 3, 2026
- Product code
- OJF
- Device type
- Hysterectomy Kit
- Reason
- The Urinary Catheter Trays contain CliniClean Pericare Wipes which may potentially be contaminated with a lubricant used during the manufacturing process. If a contaminated wipe is used on a patient, localized skin effects may occur. These may include irritation, erythema, or mild inflammation.
- Root cause
- Process change control
- Quantity
- 175 units
- Code info
- 1) PBCGLVBLI, UDI-DI: 10198956055469(each), 50198956055467(case), Lot Number: GKU1823 ; 2) PB15HYFCY, UDI-DI: 10197106474877(each), 50197106474875(case), Lot Number: NGKW2883 ; 3) PB15HYFCY, UDI-DI: 10197106474877(each), 50197106474875(case), Lot Number: NGKW2883 ; 4) PB15HYFCY, UDI-DI: 10197106474877(each), 50197106474875(case), Lot Number: NGKW2883 ; 5) SMA4FTLSCA, UDI-DI: 10197106609880(each), 50197106609888(case), Lot Number: GKU1823
- Distribution
- US Nationwide distribution.
- Voluntary or mandated
- Voluntary: Firm initiated
- Location
- Waukegan, IL, United States