Device recall
Medline Convenience Kit: OPEN HEART A PACK, Product Number DYNJ43008N
- Recall number
- Z-3137-2026
- Event id
- 99583
- Status
- Ongoing
- FDA classification
- Class II
- Recalling firm
- Medline Industries, LP
- Report date
- Sep 23, 2026
- Posted
- Sep 11, 2026
- Initiated
- Jul 23, 2026
- Center classification date
- Sep 10, 2026
- Product code
- LRO
- Device type
- General Surgery Tray
- Reason
- The kits contain a Retrograde Coronary Sinus Perfusion (RCSP) cannulae that has the potential for a sterile barrier breach. Sterility of the product cannot be assured. If a compromised pouch seal is not identified prior to use and the cannula is used, there is a potential risk of infection or other complications associated with the use of a non-sterile device, including hemolysis, foreign body reaction, thromboembolism, and/or organ dysfunction.
- Root cause
- Packaging process control
- Quantity
- 135 units
- Code info
- UDI-DI: 10198459445941 (each), 40198459445942 (case); Lot Numbers: 25HLA462, 25HLA916, 25JLA847, 25LLA298, 26CLA372
- Distribution
- US Nationwide distribution in the state of CA.
- Voluntary or mandated
- Voluntary: Firm initiated
- Related 510(k)s
- K213481
- Location
- Northfield, IL, United States