Device recall

INCA (Infant Nasal Cannulae Assy) Infant Nasal CPAP, Complete Set. Stay-Flex tubing component. 5 units per box. Model Numbers: 44-2707, 44-2709, 44-2710, 44-2712, 44-2715.

FDA recall source

Recall number
Z-3139-2026
Event id
99643
Status
Ongoing
FDA classification
Class I
Recalling firm
CooperSurgical, Inc.
Report date
Sep 30, 2026
Posted
Sep 18, 2026
Initiated
Aug 11, 2026
Center classification date
Sep 18, 2026
Product code
BZD
Device type
Ventilator, Non-Continuous (Respirator)
Reason
Potential to exhibit a loose connection between the flexible tubing and the nasal cannula, during assembly or while in use, which may result in loss of CPAP pressure, possibly reducing the effectiveness of treatment and lead to decreased oxygenation until connection is restored.
Root cause
Device Design
Quantity
2,643 boxes (13,215 units)
Code info
1. Model Number: 44-2707. UDI-DI: (01)20888937009433. Lot Code: 550019722. 2. Model Number: 44-2709. UDI-DI: (01)60888937009448. Lot Codes: 550031617, 550033781, 550031621. 3. Model Number: 44-2710. UDI-DI: (01)20888937009457. Lot Codes: 550031615, 550031616, 550031620, 550034760. 4. Model Number: 44-2712. UDI-DI: (01)20888937009464. Lot Codes: 550031611, 550031614, 550033779, 550031619. 5. Model Number: 44-2715. UDI-DI: (01)20888937009471. Lot Codes: 550031618, 550033780, 550034763.
Distribution
Nationwide distribution to CT, FL, GA, IA, IL, IN, MD, MI, MS, NJ, NY, OH, PA, VA. International distribution to Brazil and Peru.
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K911780
Location