Skip to content

Parker Quality & Compliance

PQC Reg Watch

Sponsor this week — one slot, $75, newsletter line and this bar

Device recall

IV Non-Dedicated Gravity Administration Sets with drip chambers

FDA recall source

Recall number
Z-3157-2026
Event id
99058
Status
Ongoing
FDA classification
Class II
Recalling firm
ICU Medical, Inc.
Report date
Sep 23, 2026
Posted
Sep 11, 2026
Initiated
May 12, 2026
Center classification date
Sep 11, 2026
Product code
FMG
Device type
Stopcock, I.V. Set
Reason
Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall that does not enter the fluid path, which if seen by a healthcare provider may lead to replacement of product before use, which could result in a delay in therapy.
Root cause
Process control
Quantity
2,450
Code info
All products manufactured on or before 7/4/2026. UDI-DI: 10887709079363 10887709079370 10887709079417 10887709079431 10887709079448 10887709079455 10887709097473 10887709079493 10887709079509 10887709079547
Distribution
Worldwide - US Nationwide distribution including in the states and territories of ND, TX, CT, KY, CO, AR, NC, NJ, VA, MD, OK, AK, IA, IN, KS, OR, CA, OH, WA, GA, AL, SC, MA, HI, FL, NH, NE, MN, MS, SD, MT, NY, DC, ID, WI, PA, AZ, MO, NV, TN, NM, WV, WY, IL, UT, MI, RI, VT, LA, DE, ME, VI, GU, PR and the countries of PK, PH, MY, HK, IN, DE, CR.
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K964435
Location
San Clemente, CA, United States