Oncology sets with drip chambers

FDA recall source

Recall number
Z-3158-2026
Event id
99058
Status
Ongoing
FDA classification
Class II
Recalling firm
ICU Medical, Inc.
Report date
Sep 23, 2026
Posted
Sep 11, 2026
Initiated
May 12, 2026
Center classification date
Sep 11, 2026
Product code
FPA
Device type
Set, Administration, Intravascular
Reason
Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall that does not enter the fluid path, which if seen by a healthcare provider may lead to replacement of product before use, which could result in a delay in therapy.
Root cause
Process control
Quantity
107,800
Code info
All products manufactured on or before 7/4/2026. UDI-DI: 10840619054486, 10887709091914
Distribution
Worldwide - US Nationwide distribution including in the states and territories of ND, TX, CT, KY, CO, AR, NC, NJ, VA, MD, OK, AK, IA, IN, KS, OR, CA, OH, WA, GA, AL, SC, MA, HI, FL, NH, NE, MN, MS, SD, MT, NY, DC, ID, WI, PA, AZ, MO, NV, TN, NM, WV, WY, IL, UT, MI, RI, VT, LA, DE, ME, VI, GU, PR and the countries of PK, PH, MY, HK, IN, DE, CR.
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K081361
Location
San Clemente, CA, United States