Skip to content

Parker Quality & Compliance

PQC Reg Watch

Sponsor this week — one slot, $75, newsletter line and this bar

Device recall

SunMed Tracheal Tube, Size 2.5; Model Number: 1-7333-25

FDA recall source

Recall number
Z-3172-2026
Event id
99685
Status
Ongoing
FDA classification
Class II
Recalling firm
WELL LEAD MEDICAL CO. LTD
Report date
Sep 23, 2026
Posted
Sep 15, 2026
Initiated
Jul 31, 2026
Center classification date
Sep 15, 2026
Product code
BTR
Device type
Tube, Tracheal (W/Wo Connector)
Reason
The 6Fr suction catheter could not pass through the samples smoothly.
Root cause
Process control
Quantity
27200
Code info
Model Number: 1-7333-25; UDI-DI:10889483048108; Lot No.: 2108011462, 2108011506, 2109011963, 2111012039, 2112012040, 2204010568, 2204010569, 2204010734, 2205010833, 2207011263, 2208011729, 2208011789, 2304010830, 2306011372, 2306011425, 2308011907, 2308012011, 2308012012, 2309012128, 2310012358, 2310012599, 2311013016, 2311012924, 2312013223, 2401010069, 2401010197, 2402010378, 2403010466, 2403010889, 2404010999, 2406011822, 2407012282, 2408012389, 2409012800, 2409012882, 2508013006, 2510013653, 2511014081, 2512014080, 2602010272, 2603010274, 2603010900, 2604011014, 2604011416, 2605011923, 2605011924;
Distribution
US Nationwide distribution in the state of MI.
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K042683
Location
Guangzhou, N/A, China