Brand Name: VERIQA; Product Name: RT MonteCarlo 3D v3; Model/Catalog Number: S070031.520; Software Version: 3.1.18;

FDA recall source

Recall number
Z-3174-2026
Event id
99720
Status
Ongoing
FDA classification
Class II
Recalling firm
PTW FREIBURG
Report date
Sep 23, 2026
Posted
Sep 15, 2026
Initiated
Aug 4, 2026
Center classification date
Sep 15, 2026
Product code
IYE
Device type
Accelerator, Linear, Medical
Reason
When a bolus is included in a treatment plan and sent for secondary dose calculation, the software accounts for it but does not display this clearly, which may lead users to incorrectly omit the bolus during treatment and deliver the prescribed dose to the wrong tissue depth.
Root cause
Software design
Quantity
2
Code info
Model/Catalog Number: S070031.520; UDI-DI: +EPTWS070031Q5200; Software Version 3.1.18;
Distribution
Worldwide - US Nationwide distribution in the states of GA, NY and the countries of Austria, Belgium, Switzerland, Czechia, Germany, Spain, France, Netherlands, Norway, Poland, Portugal, Romania, T¿rkiye, Chile, United Arab Emirates, Saudi Arabia, Mexico, Hong Kong, Brazil, Tajikistan, Ecuador, Iran, Thailand, Serbia, South Africa, Kazakhstan, Algeria.
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K213370
Location
Freiburg Im Breisgau, N/A, Germany