Skip to content

Parker Quality & Compliance

PQC Reg Watch

Sponsor this week — one slot, $75, newsletter line and this bar

Device recall

Kinos Axiom Total Ankle System with Axiom PSR, Flat-cut (Cut Guide), REF: 5831-9006

FDA recall source

Recall number
Z-3178-2026
Event id
99512
Status
Ongoing
FDA classification
Class II
Recalling firm
restor3d Inc.
Report date
Sep 23, 2026
Posted
Sep 16, 2026
Initiated
Jul 27, 2026
Center classification date
Sep 16, 2026
Product code
OYK
Device type
Ankle Arthroplasty Implantation System
Reason
Ankle implantation systems were packaged with incorrect patient-specific cutting guides, which may result in: delayed surgery, the need to use standard instrumentation, reduced fixation or poor alignment of implants, implant loosening/migration, or need for revision surgery.
Root cause
Labeling Change Control
Quantity
2
Code info
System Serial Numbers: 18169, 18059. REF: 5831-9006, Lot: D-788, UDI-DI: 00840097509945.
Distribution
US Nationwide distribution in the states of FL, CO.
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K223326
Location
Durham, NC, United States