Device recall

Aisys CS2 with Et Control, Datex-Ohmeda anesthesia system, Part Number 1011-9000-000

FDA recall source

Recall number
Z-3191-2026
Event id
99649
Status
Ongoing
FDA classification
Class II
Recalling firm
Datex-Ohmeda, Inc.
Report date
Sep 23, 2026
Posted
Sep 16, 2026
Initiated
Jul 30, 2026
Center classification date
Sep 16, 2026
Product code
BSZ
Device type
Gas-Machine, Anesthesia
Reason
Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect.
Root cause
Nonconforming Material/Component
Quantity
7 units
Code info
Serial Numbers: ANAS01357 ANAP00609 ANAQ00988 ANAS00586 ANAR00543 ANAP01088 ANAM00443
Distribution
Worldwide - US Nationwide distribution.
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K172045
Location
Madison, WI, United States