Device recall

Captus 4000e Thyroid Uptake System, Models: 1. 5430-30151, 2. 5430-30152, 3. 5430-30154;

FDA recall source

Recall number
Z-3193-2026
Event id
99677
Status
Ongoing
FDA classification
Class II
Recalling firm
Mirion Technologies (Capintec), Inc.
Report date
Sep 23, 2026
Posted
Sep 16, 2026
Initiated
Jul 20, 2026
Center classification date
Sep 16, 2026
Product code
IZD
Device type
Probe, Uptake, Nuclear
Reason
Potential for failure of the tension bolt which would allow the spring arm to unexpectedly fall to its lowest position, which is approximately 25 inches (63.5 cm) from the ground.
Root cause
Device Design
Quantity
783
Code info
Affected Units: Captus 4000: 1. Model 5430-30151; UDI-DI: 00859942006096; 2. Model 5430-30152; UDI-DI: 00859942006102; 3. Model 5430-30154; UDI-DI: 00859942006423; Devices shipped from April 2019 through February 2024, Serial Numbers 940548 through 941398, that have NOT already replaced the original arm with the redesigned arm.
Distribution
Worldwide distribution - US Nationwide and the countries of ALGERIA, AUSTRIA, BANGLADESH, BRASIL, Brazil, CANADA, CHILE, Costa Rica, Ecuador, France, Hanoi Vietnam, Hong Kong, INDIA, ISRAEL, ITALY, Malaysia, Philippines, POLAND, Saudi Arabia, SOUTH KOREA, Spain, SWEDEN, TAIWAN, THAILAND, TURKEY, UNITED ARAB EMIATES, UNITED KINGDOM, VIETNAM.
Voluntary or mandated
Voluntary: Firm initiated
Location
Florham Park, NJ, United States