TRIATHLON HINGE INSERT SIZE 4 11MM 5612-P-411 TRIATHLON HINGE INSERT SIZE 4 13MM 5612-P-413 Orthopaedic implant components manufactured from UHMWPE (polyethylene)

FDA recall source

Recall number
Z-3198-2026
Event id
99558
Status
Ongoing
FDA classification
Class II
Recalling firm
Howmedica Osteonics Corp.
Report date
Sep 23, 2026
Posted
Sep 16, 2026
Initiated
Jul 30, 2026
Center classification date
Sep 16, 2026
Product code
KRO
Device type
Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer
Reason
Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
Root cause
Process change control
Quantity
12,156 US
Code info
5612-P-411 UDI-DI 07613327601893 Lot 360958 5612-P-413 UDI-DI 07613327602012 Lot 354573
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa.
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K223528
Location
Mahwah, NJ, United States