Device recall
BEARING AND AXLE BUSHING ASSY STD ELBOW C-M113-1-301 KRH REV TIB INS 16mm THK .640 DIA STEM C-M113-5-600 SELF-PRESSURIZING GLENOID SIZE 9 C-M063-5-800 Orthopaedic implant components manufactured from UHMWPE (polyethylene)
- Recall number
- Z-3202-2026
- Event id
- 99558
- Status
- Ongoing
- FDA classification
- Class II
- Recalling firm
- Howmedica Osteonics Corp.
- Report date
- Sep 23, 2026
- Posted
- Sep 16, 2026
- Initiated
- Jul 30, 2026
- Center classification date
- Sep 16, 2026
- Product code
- LGE
- Device type
- Prosthesis, Knee, Femorotibial, Semi-Constrained, Cemented, Trunnion-Bearing
- Reason
- Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
- Root cause
- Process change control
- Quantity
- 12,156 US
- Code info
- C-M113-1-301 UDI-DI 07613327590074 Lot MH621A C-M113-5-600 No UDI Lot 194593 C-M063-5-800 No UDI Lot RX06358A
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa.
- Voluntary or mandated
- Voluntary: Firm initiated
- Location
- Mahwah, NJ, United States