Device recall
SCORPIO RECESSED METAL BACKED PATELLA 3045-0026 3045-0028 3045-0030 3045-0032 Orthopaedic implant components manufactured from UHMWPE (polyethylene)
- Recall number
- Z-3203-2026
- Event id
- 99558
- Status
- Ongoing
- FDA classification
- Class II
- Recalling firm
- Howmedica Osteonics Corp.
- Report date
- Sep 23, 2026
- Posted
- Sep 16, 2026
- Initiated
- Jul 30, 2026
- Center classification date
- Sep 16, 2026
- Product code
- MBH
- Device type
- Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer
- Reason
- Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
- Root cause
- Process change control
- Quantity
- 12,156 US
- Code info
- 3045-0026 No UDI Lots 12WAHA 12WAHB HOHE HOVX 3045-0028 No UDI Lots 13VAW 13VAWA 3045-0030 No UDI Lots BIYIE HNZY 3045-0032 No UDI Lots BIYIH HOYK
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa.
- Voluntary or mandated
- Voluntary: Firm initiated
- Related 510(k)s
- K033972
- Location
- Mahwah, NJ, United States