Skip to content

Parker Quality & Compliance

PQC Reg Watch

Sponsor this week — one slot, $75, newsletter line and this bar

Device recall

SCORPIO + MOBILE BEARING PCS 78-5-0910 SYSTEM ONE BIPOLAR 40MM 6074-4022 SYSTEM ONE BIPOLAR 43MM 6074-4322 SYSTEM ONE BIPOLAR 44MM 6074-4422 Orthopaedic implant components manufactured from UHMWPE (polyethylene)

FDA recall source

Recall number
Z-3205-2026
Event id
99558
Status
Ongoing
FDA classification
Class II
Recalling firm
Howmedica Osteonics Corp.
Report date
Sep 23, 2026
Posted
Sep 16, 2026
Initiated
Jul 30, 2026
Center classification date
Sep 16, 2026
Reason
Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
Root cause
Process change control
Quantity
12,156 US
Code info
78-5-0910 UDI-DI 07613327366006 Lot 30582401 6074-4022 UDI-DI 07613327031775 Lots 2M5NTY LH25R5 M36N83 6074-4322 UDI-DI 07613327031805 Lot W40E3T 6074-4422 UDI-DI 07613327031812 Lot R84K20
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa.
Voluntary or mandated
Voluntary: Firm initiated
Location
Mahwah, NJ, United States