CADD Yellow Medication Cassette Reservoir with NRFit connector US Product Codes: 21-7600-24, 21-7609-24 OUS Product Codes: 21-7609-24 21-7600-24 21-7600-24 21-7600-24 21-7609-24 21-7609-24 21-7600-24 21-7600-24 21-7609-24 21-7609-24 21-7600-24 21-7600-24 21-7609-24 21-7600-24 21-7600-24 21-7609-24 21-7609-24JP 21-7600-24JP 21-7609-24 21-7609-24 21-7600-24 21-7600-24 21-7609-24 21-7600-24 21-7600-24 21-7600-24 21-7609-24

FDA recall source

Recall number
Z-3212-2026
Event id
99648
Status
Ongoing
FDA classification
Class II
Recalling firm
ICU Medical, Inc.
Report date
Sep 23, 2026
Posted
Sep 17, 2026
Initiated
Aug 20, 2026
Center classification date
Sep 17, 2026
Product code
LHI
Device type
Set, I.V. Fluid Transfer
Reason
Reservoirs may exhibit medication leakage at the joint between luer connector and tubing. Issue could result in delay or interruption of therapy, under delivery of medication, exposure to infectious or toxic agents, or air embolus.
Root cause
Process control
Quantity
185,688 units
Code info
21-7600-24 UDI-DI 30610586044001 Lots 6026878 6037784 6054015 6062724 6070616 6082332 6092886 6101906 6108556 6126029 6140893 6143454 6147696 6147697 6162769 6155095 6166032 6171587 6171586 21-7609-24 UDI-DI 30610586044018 Lot 6022079 6037697 6037698 6053978 6061525 6072970 6072964 6070575 6077773 6092846 6101798 6108514 6108515 6108516 6140862 6140861 6140865 6155132
Distribution
Worldwide - US Nationwide distribution in the states of AR, CA, FL, IL, IN, LA, MI, MO, MS, TN, TX, VA, WI and the countries of Austria, Australia, Belgium, Costa Rica, Germany, Estonia, Spain, Finland, France, Great Britain, Ireland, Italy, Japan, Kuwait, Netherlands, Norway, and Sweden.
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K162219
Location
Lake Forest, IL, United States