Skip to content

Parker Quality & Compliance

PQC Reg Watch

Sponsor this week — one slot, $75, newsletter line and this bar

Device recall

Brand Name: Heidelberg Engineering GmbH Product Name: SPECTRALIS with Flex Module Model/Catalog Number: S3610, S2610, S2000, S1610 multimodal imaging platform optimized for the posterior segment of the eye

FDA recall source

Recall number
Z-3215-2026
Event id
99595
Status
Ongoing
FDA classification
Class II
Recalling firm
Heidelberg Engineering GmbH
Report date
Sep 30, 2026
Posted
Sep 18, 2026
Initiated
Aug 6, 2026
Center classification date
Sep 18, 2026
Product code
OBO
Device type
Tomography, Optical Coherence
Reason
Wheels may come loose when module is being pushed or pulled during movement or transportation, affecting stability of the stand. The stand may tip over as a result. Falling stand may cause impact or crushing injuries, particularly to the feet, legs, hands or arms of the operator or nearby persons.
Root cause
Component design/selection
Quantity
87
Code info
UDI-DI 04059237000235 Lot Code: Spec-07265, Spec-11121, Spec-11281, Spec-12096, Spec-13305, Spec-14403, Spec-15282, Spec-15587, Spec-20071, Spec-20348, Spec-20351, Spec-20591, Spec-20605, Spec-23473, Spec-23941, Spec-23950, Spec-23975, Spec-24507, Spec-24558, Spec-24582, Spec-24611
Distribution
Worldwide - US Nationwide distribution in the state of Massachusetts and the countries of Australia, Costa Rica, Germany, France, India, Israel, Italy, Canada, Mexico, Netherlands, Austria, Philippines, Poland, Saudi Arabia, Switzerland, Spain. South Africa. Thailand, Czech Republic, United Kingdom, Vietnam.
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K181594
Location
Heidelberg, Germany