Device recall
Olympus EZDilate Endoscopic Balloon Dilator Wired-guided 11, 12, 13mm. Model Number: BD-420X-1355.
- Recall number
- Z-3216-2026
- Event id
- 99627
- Status
- Ongoing
- FDA classification
- Class II
- Recalling firm
- Olympus Corporation of the Americas
- Report date
- Sep 30, 2026
- Posted
- Sep 18, 2026
- Initiated
- Aug 19, 2026
- Center classification date
- Sep 18, 2026
- Product code
- FGE
- Device type
- Stents, Drains And Dilators For The Biliary Ducts
- Reason
- Potential for the inflation chart label affixed to the shaft of the device to be incorrect and indicate a balloon diameter of 18, 19, or 20 mm,
- Root cause
- Process control
- Quantity
- 305 units
- Code info
- Model Number: BD-420X-1355. UDI-DI: 00821925042261. Lot Number: 460898.
- Distribution
- International distribution in the country of Japan.
- Voluntary or mandated
- Voluntary: Firm initiated
- Location
- Center Valley, PA, United States