Device recall
Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories, Hand Pendants: MDR510094 PENDANT F/ MDR107002E MDR520094 PENDANT F/ MDR107003E MDR530094 PENDANT, W/ 9V BATTERY BACKUP, F/ 2E BED MDR540094 PENDANT, W/ 9V BATTERY BACKUP, F/ 3E BED
- Recall number
- Z-3227-2026
- Event id
- 99827
- Status
- Ongoing
- FDA classification
- Class I
- Recalling firm
- Medline Industries, LP
- Report date
- Oct 7, 2026
- Posted
- Oct 1, 2026
- Initiated
- Aug 14, 2026
- Center classification date
- Sep 30, 2026
- Product code
- LLI
- Device type
- Bed, Therapeutic, Ac-Powered, Adjustable Home-Use
- Reason
- Potential fire hazard involving the hand control pendant and associated wiring used with affected beds. Pendant or pendant wiring may overheat, potentially leading to smoking, sparking, melting, burning, or fire. Additional products added to 11/26/2025 recall.
- Root cause
- Device Design
- Quantity
- 771,127 total
- Code info
- MDR510094 UDI-DI 10888277363021 MDR520094 UDI-DI 10884389161692 MDR530094 UDI-DI 10197344119059 MDR540094 UDI-DI 10197344119066 All lots
- Distribution
- Distribution US Nationwide including territories of Puerto Rico and Guam and the countries of Bermuda, Canada, and Panama.
- Voluntary or mandated
- Voluntary: Firm initiated
- Location
- Northfield, IL, United States