Device recall
Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories MDR107000P BED, MANUAL, MEDLITE
- Recall number
- Z-3228-2026
- Event id
- 99827
- Status
- Ongoing
- FDA classification
- Class I
- Recalling firm
- Medline Industries, LP
- Report date
- Oct 7, 2026
- Posted
- Oct 1, 2026
- Initiated
- Aug 14, 2026
- Center classification date
- Sep 30, 2026
- Product code
- FNJ
- Device type
- Bed, Manual
- Reason
- Potential fire hazard involving the hand control pendant and associated wiring used with affected beds. Pendant or pendant wiring may overheat, potentially leading to smoking, sparking, melting, burning, or fire. Additional products added to 11/26/2025 recall.
- Root cause
- Device Design
- Quantity
- 771,127 total
- Code info
- MDR107000P UDI-DI 00080196288284 All lots
- Distribution
- Distribution US Nationwide including territories of Puerto Rico and Guam and the countries of Bermuda, Canada, and Panama.
- Voluntary or mandated
- Voluntary: Firm initiated
- Location
- Northfield, IL, United States