Device recall

VITEK 2 AST-N461 TEST KIT, GRAM NEGATIVE SUSCEPTIBILITY TESTING - AUTOMATED, Reference Number 424811

FDA recall source

Recall number
Z-3231-2026
Event id
99749
Status
Ongoing
FDA classification
Class II
Recalling firm
Biomerieux Inc
Report date
Sep 30, 2026
Posted
Sep 18, 2026
Initiated
Jul 24, 2026
Center classification date
Sep 18, 2026
Product code
LON
Device type
System, Test, Automated, Antimicrobial Susceptibility, Short Incubation
Reason
The issue is related to card pouch damage associated with a portion of the lot which was produced on a specific card pouching device, in a specific timeframe, causing a potential risk of false resistant results. False resistant results may lead to the use of less effective antibiotic treatments or the association of multiple antibiotics having negative side effects, which could potentially impact patient s health.
Root cause
Packaging process control
Quantity
97680 units
Code info
UDI-DI 03573026637712; Lot Number 1333393504
Distribution
International distribution to the countries of Austria, Germany, Luxembourg.
Voluntary or mandated
Voluntary: Firm initiated
Location
Hazelwood, MO, United States