Device recall

Name (Model Number): 1. Intera 1.5T (781195), 2. Intera 1.5T (781295), 3. Intera 1.5T Achieva IT Nova (781175), 4. Intera 1.5T Achieva Nova (781172), 5. Intera 1.5T Achieva Nova-Dual (781173), 6. Intera 1.5T Omni/Stellar (781104), 7. Intera Achieva 1.5T Pulsar (781171);

FDA recall source

Recall number
Z-3240-2026
Event id
99769
Status
Ongoing
FDA classification
Class II
Recalling firm
Philips North America Llc
Report date
Sep 30, 2026
Posted
Sep 18, 2026
Initiated
Aug 21, 2026
Center classification date
Sep 18, 2026
Product code
LNH
Device type
System, Nuclear Magnetic Resonance Imaging
Reason
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
Root cause
Software design
Quantity
82
Code info
1. Intera 1.5T Model Numbers (REF): 781195; UDI-DI: N/A; Serial Numbers: 18774, 18781, 18951, 18958, 18962, 18971, 20431, 18898, 20429, 18669, 18981, 18982, 18734, 18763, 20427, 18842, 18654, 18878, 18566, 18618, 18752, 20409, 18748, 18743, 18657, 20424, 18755, 18908, 18501, 18646, 18768, 18516, 18554, 18616, 18723, 18725, 18784; 2. Intera 1.5T; Model Numbers (REF): 781295; UDI-DI: N/A; Serial Numbers: 30138, 30297, 30314, 30210, 30137, 30173, 30179, 30104, 30105, 30290, 30208, 30051, 30035, 30143, 30109, 30063, 30237, 30135; 3. Intera 1.5T Achieva IT Nova; Model Numbers (REF): (781175); UDI-DI: N/A; Serial Number: 15003; 4. Intera 1.5T Achieva Nova; Model Numbers (REF): 781172; UDI-DI: N/A; Serial Numbers: 12105, 12049, 12079, 12001, 12086, 12099, 12046, 12057, 12081, 12075, 12078, 12106; 5. Intera 1.5T Achieva Nova-Dual; Model Numbers (REF): 781173; UDI-DI: N/A; Serial Numbers: 13006, 13017, 13058, 13012, 13048, 13049; 6. Intera 1.5T Omni/Stellar; Model Numbers (REF): 781104; UDI-DI: N/A; Serial Number: 10403; 7. Intera Achieva 1.5T Pulsar; Model Numbers (REF): 781171; UDI-DI: 884838088108; Serial Numbers: 11044, 11009, 11029, 11043, 11040, 11039, 11090;
Distribution
Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Andorra, Angola, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Belize, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Equatorial Guinea, Estonia, Fiji, Finland, France, French Polynesia, Georgia, Germany, Ghana, Greece, Guadeloupe, Guernsey, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macao, Macedonia, Madagascar, Malaysia, Maldives, Malta, Martinique, Mauritius, Mayotte, Mexico, Moldova, Monaco, Mongolia, Montenegro, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, State of, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Viet Nam, Yemen, Zambia;
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K030520K052013K052078
Location
Cambridge, MA, United States