Device recall

nGenuity Patient Monitor with Capnography (ETCO2) CAT 8100E1-ST-US, CAT 8100E1-US, CAT 8100EP1, CAT 8100EP1-ST, CAT 8100EP1-ST-US, CAT 8100EP1-US, CAT 8100EP1R

FDA recall source

Recall number
Z-3253-2026
Event id
99545
Status
Ongoing
FDA classification
Class II
Recalling firm
Criticare Technologies, Inc.
Report date
Sep 30, 2026
Posted
Sep 22, 2026
Initiated
May 6, 2026
Center classification date
Sep 22, 2026
Product code
MHX
Device type
Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Reason
Complaint trend where "Checkline Error" alarm presents at device setup, rendering the CO2 monitoring function inoperable. May result in procedure delay.
Root cause
Process design
Quantity
1,007
Code info
UDI-DI +B4388100EP1US02 All units with CO2 option SN beginning with 424, 125, 225, 325, 425, 126, 226, 326, 426, date range 2024-1st qt 2026
Distribution
US Nationwide distribution including in the states and territory of AL, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, ND, NH, NJ, NV, NY, OH, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and Puerto Rico.
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K030613
Location
Warwick, RI, United States