Skip to content

Parker Quality & Compliance

PQC Reg Watch

Sponsor this week — one slot, $75, newsletter line and this bar

Device recall

PRO+ MRS SURFACE FOR VERSACARE, mattress with Advanced Microclimate Technology (AMT) uses a non-powered system of intake and output valves to provide weight-based pressure redistribution whenever a patient moves or is repositioned, Model/Catalog Number: P3255A08

FDA recall source

Recall number
Z-3262-2026
Event id
99606
Status
Ongoing
FDA classification
Class II
Recalling firm
Baxter Healthcare Corporation
Report date
Sep 30, 2026
Posted
Sep 24, 2026
Initiated
Aug 18, 2026
Center classification date
Sep 24, 2026
Product code
IOQ
Device type
Bed, Flotation Therapy, Powered
Reason
There is an error in the product description on the Unique Device Identification (UDI) label. These discrepancies in labeling do not align with the approved product or product specifications and may lead to incorrect product identification or selection, potentially resulting in use with an incompatible bed frame.
Root cause
Process control
Quantity
169 units
Code info
Serial Number: A171BP3769 - B057BP6765 UDI-DI Number: 00887761007543
Distribution
Worldwide distribution - US Nationwide and the country of Canada.
Voluntary or mandated
Voluntary: Firm initiated
Location
Deerfield, IL, United States