Skip to content

Parker Quality & Compliance

PQC Reg Watch

Sponsor this week — one slot, $75, newsletter line and this bar

Device recall

McKesson CONSULT hCG Urine Cassette, Model/ Part: FHC-A102/5001; Alere hCG Cassette (20 mIU/mL) in Urine, Model/ Part: FHC-A102/92210; Fisher Scientific Sure-Vue Serum/Urine¿hCG-STAT, Model/ Part: FHC-A202/23900531, FHC-A202/23900530; McKesson CONSULT Diagnostics hCG Urine Cassette, Model/ Part: FHC-A202/5002; Alere hCG Combo Cassette (20/10 mIU/mL) for Urine or Serum, Model/ Part: FHC-A202/92215; Cardinal Health hCG Combo Rapid Test Serum or Urine, Model/ Part: FHC-A202/B-1077-23; Medline hCG Combo Pregnancy Test Cassette, Model/ Part: FHC-A202/MPH22025HCG;

FDA recall source

Recall number
Z-3271-2026
Event id
99409
Status
Ongoing
FDA classification
Class II
Recalling firm
Alere San Diego, Inc.
Report date
Oct 7, 2026
Posted
Sep 25, 2026
Initiated
Aug 20, 2026
Center classification date
Sep 25, 2026
Product code
JHI
Device type
Visual, Pregnancy Hcg, Prescription Use
Reason
hCG rapid tests may detect low levels of hCG below the stated assay sensitivity, which may result in false positive result.
Root cause
Device Design
Quantity
50,265,835
Code info
All lots within the expiration date. Parts/UDI-DI/Labeling Versions Through: 5001/ 20612479202744/1155958203, 12/15; 92210/10815845020130/1156015803, 07/16; 23900530/00613647000571/1155810105, 2018-01-15; 23900531/00613647000533/1155810105, 2018-01-15; 5002/20612479202768/1155958305, 12/15; 92215/10815845020123/1156015903, 07/16; B-1077-23/20885380020268/1155904805, 9/2014; MPH22025HCG/10884389191880/1156093402, 04-01-2015
Distribution
Worldwide - US Nationwide distribution including in the states of IL, PA, FL, GA, VA, NY, TN, AZ, ID, ND, WI, AL, LA, NJ,MT, NE, OH, PR, IN, MI, WY, MA, CA, TX, HI, KS, KY, AR, CO, NC, WA, IA, MN, OK, OR, NM, SC, GU, MD, NV, MS, WV, SD, VT, DE and the countries of United Kingdom, Canada, Italy, Ireland, Spain, Portugal, New Zealand, Hong Kong, Australia, Papua New Guinea, Fiji, Guam, Sweden.
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K062361
Location
San Diego, CA, United States