Skip to content

Parker Quality & Compliance

PQC Reg Watch

Sponsor this week — one slot, $75, newsletter line and this bar

Device recall

Fisher Scientific Sure-Vue Serum/Urine hCG, Model/ Part: FHC-202/23900529, FHC-202/23900528; Beckman Coulter Inc's ICON 25 hCG, Model/ Part: FHC-202/43025; Henry Schein One Step+ hCG Combo test in Serum or Urine, Model/ Part: FHC-202/9004075; Alere hCG Combo Cassette (25mIU/mL) in Urine (CLIA Waived) or Serum, Model/ Part: FHC-202/92225; Pro Advantage By NDC Urine/Serum Pregnancy Cassette Device, Model/ Part: FHC-202/P080021

FDA recall source

Recall number
Z-3272-2026
Event id
99409
Status
Ongoing
FDA classification
Class II
Recalling firm
Alere San Diego, Inc.
Report date
Oct 7, 2026
Posted
Sep 25, 2026
Initiated
Aug 20, 2026
Center classification date
Sep 25, 2026
Product code
JHI
Device type
Visual, Pregnancy Hcg, Prescription Use
Reason
hCG rapid tests may detect low levels of hCG below the stated assay sensitivity, which may result in false positive result.
Root cause
Device Design
Quantity
5,331,585
Code info
All lots within the expiration date. Parts/UDI-DI/Labeling Versions Through: 23900529/00613647000540/1155810906, 2018-01-15; 23900528/00613647000564/1155810906, 2018-01-15; 43025/15099590776886/May 2015; 9004075/10304040024123/1155875804, 2014-06-09; 92225/10815845020116/1156015603, 07/16; P080021/M220P0800213/40001856, 2023-03-30
Distribution
Worldwide - US Nationwide distribution including in the states of IL, PA, FL, GA, VA, NY, TN, AZ, ID, ND, WI, AL, LA, NJ,MT, NE, OH, PR, IN, MI, WY, MA, CA, TX, HI, KS, KY, AR, CO, NC, WA, IA, MN, OK, OR, NM, SC, GU, MD, NV, MS, WV, SD, VT, DE and the countries of United Kingdom, Canada, Italy, Ireland, Spain, Portugal, New Zealand, Hong Kong, Australia, Papua New Guinea, Fiji, Guam, Sweden.
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K993065
Location
San Diego, CA, United States