Skip to content

Parker Quality & Compliance

PQC Reg Watch

Sponsor this week — one slot, $75, newsletter line and this bar

Device recall

Brand Name: Opportunity Pro Product Name: Opportunity Pro Model/Catalog Number: 72-50800, 72-50801,72-50802,72-50803,72-50804 Product Description: Blue Nitrile patient examination glove

FDA recall source

Recall number
Z-3291-2026
Event id
99748
Status
Ongoing
FDA classification
Class II
Recalling firm
American Nitrile Operations Llc
Report date
Oct 7, 2026
Posted
Sep 25, 2026
Initiated
Aug 14, 2026
Center classification date
Sep 25, 2026
Product code
LZA
Device type
Polymer Patient Examination Glove
Reason
Gloves may exhibit defects such as tears, missing fingers, and debris. Use of defective gloves may lead to healthcare personnel and patient exposure to infectious pathogens leading to infection or cross-contamination, or exposure to chemical agents leading to irritation or systemic toxicities.
Root cause
Process control
Quantity
17,481,000
Code info
Item No. 72-RM50800 UDI-DI 818634025487 Lots AO01519 Item No. 72-RM50801 UDI-DI 818634025401 Lots AO01518 AO01564 Item No. 72-RM50802 UDI-DI 818634025418 Lots AO01517 AO01531 AO01530 AO01568 Item No. 72-RM50803 UDI-DI 818634025425 Lots AO01529 AO01567 AO01593 AO01594 Item No. 72-RM50804 UDI-DI 818634025432 Lots AO01222 AO01528 AO01566 Alternate lots AN250310948, AN250310947, AN250310945, AN250311327
Distribution
US Nationwide distribution.
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K220825
Location
Grove City, OH, United States