Skip to content

Parker Quality & Compliance

PQC Reg Watch

Sponsor this week — one slot, $75, newsletter line and this bar

Device recall

Presource Packs Containing Stryker StrykeFlow PS and AHTO tube sets: 1) KIT,GU DAVINCI , Catalog Number: PGVMDVBWU; 2) KIT,GU DAVINCI , Catalog Number: PGVMDVBWV; 3) KIT,GU DAVINCI , Catalog Number: PGVMDVB26; 4) KIT,GU DAVINCI , Catalog Number: PGVMDVB27; 5) KIT,GU DAVINCI , Catalog Number: PGVMDVB28; 6) KIT,GU DAVINCI , Catalog Number: PGVMDVB29; 7) COLORECTAL PACK , Catalog Number: SBA73CLKFD; 8) COLORECTAL PACK , Catalog Number: SBA73CLK26

FDA recall source

Recall number
Z-3296-2026
Event id
99768
Status
Ongoing
FDA classification
Class II
Recalling firm
Cardinal Health 200, LLC
Report date
Oct 7, 2026
Posted
Sep 25, 2026
Initiated
Aug 18, 2026
Center classification date
Sep 25, 2026
Product code
NWU
Device type
Endoscope Introducer Kit
Reason
The kits contain tube sets with handpieces that may leak during use. The impacted devices can have the following failures: fluid leakage from the hand piece that can pool on the floor. The impact can range from user dissatisfaction to the potential for the user of slip and fall.
Root cause
Under Investigation by firm
Quantity
344 kits
Code info
1) PGVMDVBWU, UDI-DI: 10888439626186(each), 50888439626184(case), Lot Number: 24317FG2 ; 2) PGVMDVBWU, UDI-DI: 10888439626186(each), 50888439626184(case), Lot Number: 24353FG2 ; 3) PGVMDVBWU, UDI-DI: 10888439626186(each), 50888439626184(case), Lot Number: 24317FG2 ; 4) PGVMDVBWU, UDI-DI: 10888439626186(each), 50888439626184(case), Lot Number: 25069FG2 ; 5) PGVMDVBWU, UDI-DI: 10888439626186(each), 50888439626184(case), Lot Number: 25045FG2 ; 6) PGVMDVBWU, UDI-DI: 10888439626186(each), 50888439626184(case), Lot Number: 25069FG2 ; 7) PGVMDVBWV, UDI-DI: 10198956204034(each), 50198956204032(case), Lot Number: 24353FG2 ; 8) PGVMDVBWV, UDI-DI: 10198956204034(each), 50198956204032(case), Lot Number: 25101FG2 ; 9) PGVMDVBWV, UDI-DI: 10198956204034(each), 50198956204032(case), Lot Number: 25148FG2 ; 10) PGVMDVBWV, UDI-DI: 10198956204034(each), 50198956204032(case), Lot Number: 25217FG2 ; 11) PGVMDVB26, UDI-DI: 10198956007796(each), 50198956007794(case), Lot Number: 24309FG2 ; 12) PGVMDVB27, UDI-DI: 10198956054714(each), 50198956054712(case), Lot Number: 24325FG2 ; 13) PGVMDVB28, UDI-DI: 10198956079847(each), 50198956079845(case), Lot Number: 24317FG2 ; 14) PGVMDVB28, UDI-DI: 10198956079847(each), 50198956079845(case), Lot Number: 24325FG2 ; 15) PGVMDVB29, UDI-DI: 10198956094024(each), 50198956094022(case), Lot Number: 24317FG2 ; 16) PGVMDVB29, UDI-DI: 10198956094024(each), 50198956094022(case), Lot Number: 24348FG2 ; 17) SBA73CLKFD, UDI-DI: 10197106633960(each), 50197106633968(case), Lot Number: 25055FG2 ; 18) SBA73CLK26, UDI-DI: 10198956029828(each), 50198956029826(case), Lot Number: 24309FG2
Distribution
US Nationwide distribution in the states of CA, IL, KY, MO, NC, NY, OH, TX.
Voluntary or mandated
Voluntary: Firm initiated
Location
Waukegan, IL, United States