Skip to content

Parker Quality & Compliance

PQC Reg Watch

Sponsor this week — one slot, $75, newsletter line and this bar

Device recall

Brand Name: Tandem t:slim mobile application Product Name: t:slim Mobile Application, iOS Model/Catalog Number: iOS software version 2.13 Software Version: iOS software version 2.13 UDI-DI: 00389152470740 Component: No

FDA recall source

Recall number
Z-3330-2026
Event id
99589
Status
Ongoing
FDA classification
Class II
Recalling firm
Tandem Diabetes Care, Inc.
Report date
Oct 7, 2026
Posted
Sep 29, 2026
Initiated
Jul 16, 2026
Center classification date
Sep 29, 2026
Product code
QFG
Device type
Alternate Controller Enabled Insulin Infusion Pump
Reason
A beta (early-release) version of the iOS application, version 2.13 (TestFlight) had a software issue preventing successful pairing of the t:slim application with Abbott FreeStyle Libre 3 Plus (FSL3P) continuous glucose monitoring (CGM) sensors, resulting in the insulin pump being unable to establish a near-field communication (NFC) connection with the CGM and requiring FSL3P users to use their t:slim insulin pumps in manual mode. FSL3P users are unable to access their CGM's glucose values from within the t:slim app and cannot use their pumps in automated control (closed-loop) mode, which could lead to users experiencing hypoglycemia or hyperglycemia.
Root cause
Software Design Change
Quantity
177
Code info
Lot Code: UDI-DI: 00389152470641 Software Version: 2.13 (iOS only)
Distribution
US Nationwide distribution.
Voluntary or mandated
Voluntary: Firm initiated
Related 510(k)s
K201214K203234K232380
Location
San Diego, CA, United States